Understanding Charge Amplifiers
Definition: What is a Charge Amplifier?
Charge amplifier is an electronic signal conditioning device that converts the high-impedance charge output (measured in picocoulombs, pC) from charge-mode piezoelectric accelerometers into a low-impedance voltage output suitable for transmission over cables and processing by measurement instruments. The charge amplifier acts as an impedance converter and amplifier, enabling the use of charge-mode sensors which can operate at extreme temperatures and harsh conditions where IEPE accelerometers would fail.
While less common in routine industrial monitoring (replaced by simpler IEPE sensors), charge amplifiers remain essential for specialized applications requiring extreme temperature capability (above 175°C), nuclear environments, or situations where sensor electronics cannot be tolerated. Understanding charge amplifier operation is important for high-temperature vibration monitoring and historical measurement systems.
Operating Principle
Charge-to-Voltage Conversion
- Piezoelectric sensor generates charge (Q) proportional to acceleration
- Charge collected on special low-noise cable capacitance
- Charge amplifier integrates charge using feedback capacitor
- Output voltage V = Q / Cfeedback
- Result: Low-impedance voltage output (typically ±10V full scale)
Key Circuit Features
- Very high input impedance (>10^12 ohms) to avoid charge leakage
- Feedback capacitor defines gain/sensitivity
- Feedback resistor sets low-frequency response
- Low-noise design critical for weak signals
- Multiple gain settings for different sensor sensitivities
Advantages of Charge Mode Systems
Extreme Temperature Capability
- Charge-mode sensors operate to 650°C (some to 1000°C)
- No electronics in sensor to fail from heat
- Essential for exhaust systems, furnaces, engines
- IEPE limited to ~175°C maximum
Radiation Resistance
- No active electronics in sensor
- Suitable for nuclear environments
- IEPE electronics damaged by radiation
Cable Interchangeability
- Can change cable length without recalibration
- Charge insensitive to cable capacitance (within limits)
- Flexibility in installation
Disadvantages and Challenges
System Complexity
- Requires separate external charge amplifier (cost, size)
- More components = more potential failure points
- Setup and configuration more complex than IEPE
Cable Requirements
- Must use special low-noise cable
- Cable motion can generate noise (triboelectric effect)
- Cable must be secured to prevent vibration
- More expensive than standard coax
- Practical length limit ~100m typically
Sensitivity to Moisture
- High impedance sensitive to insulation resistance
- Moisture can cause signal drift or noise
- Requires good sealing and cable condition
When to Use Charge Mode
Required Applications
- High Temperature: >175°C (exhaust systems, furnaces, kilns, engine testing)
- Nuclear Environments: Radiation exceeding electronics tolerance
- Explosive Atmospheres: Intrinsically safe sensors without active electronics
- Research: Specialized testing requiring charge-mode characteristics
Not Recommended When
- Standard industrial monitoring (use IEPE instead)
- Long cable runs in electrically noisy environment
- Budget constraints (charge amplifiers expensive)
- Routine condition monitoring (complexity not justified)
Charge Amplifier Features
Gain/Sensitivity Settings
- Adjustable to match sensor sensitivity
- Typical ranges: 0.1-1000 mV/pC
- Allows using different sensors with same amplifier
- Must be calibrated for sensor being used
Frequency Response Control
- High-pass filter cutoff adjustable (0.1-10 Hz typical)
- Low-pass filter for anti-aliasing
- Integration/differentiation functions
- Optimized for application requirements
Cable Drive Capability
- Low-impedance output drives long cables to instruments
- Typically ±10V output
- Can drive multiple instruments if needed
Setup and Calibration
Configuration
- Connect sensor to charge amplifier with low-noise cable
- Set amplifier gain to match sensor sensitivity
- Set frequency range (high-pass and low-pass filters)
- Connect amplifier output to measurement instrument
- Verify end-to-end calibration with known excitation
Calibration Verification
- Shaker table calibration
- Portable calibrator (handheld exciter)
- Back-to-back comparison with reference sensor
- Check sensitivity and frequency response
Modern Trends
Declining Use
- IEPE has replaced charge mode in most applications
- Simpler, lower cost, easier to use
- Charge mode relegated to specialized applications
- Some facilities phasing out charge-mode systems
Remaining Applications
- High-temperature monitoring (gas turbines, engines)
- Nuclear power plants
- Research laboratories
- Precision measurements requiring charge-mode advantages
- Legacy systems maintenance
Charge amplifiers are specialized signal conditioning devices that enable use of charge-mode piezoelectric accelerometers in extreme conditions where IEPE sensors cannot operate. While their complexity and cost have limited them to specialized applications, understanding charge amplifier operation remains important for high-temperature vibration monitoring and maintaining legacy measurement systems in industrial facilities.