Meet Vibromera.com — our new international website. Visit Vibromera.com →
Traceable coherent-vector arithmetic

Coherent Phasor Sum and Difference Worksheet

Combine two documented same-frequency phasors, preserve the signed phase relation and choose the subtraction order explicitly. This worksheet does not diagnose a fault or calculate a balancing correction mass or angle.

Signed relative phaseExplicit subtraction orderZero-vector phase undefinedNo automatic diagnosis
Phasors are comparable only under one conventionUse the same physical quantity, unit, amplitude convention, coherent frequency/order, reference event, phase sign convention, sensor polarity and operating condition. A phase number copied from another analyzer mode can mean something different.

Measurement identity

Blank fields are deliberate. The worksheet cannot infer whether an entered phase is lead or lag, peak- or zero-crossing-referenced, or relative to a keyphasor or another order.

Two coherent phasors

Finite and positive, in the declared common unit/scaling.
Any finite angle; normalized for display.
Evidence gates

The arithmetic can be displayed with incomplete gates, but it must not be treated as a reviewed physical vector comparison until all six are documented.

Δεν έχει υπολογιστεί

    Signed relative phase, V2 minus V1
    Smallest unsigned separation
    Vector sum
    Selected vector difference
    Normalized V1 phase
    Normalized V2 phase
    Vectorx componenty componentΠλάτοςΦάση
    V1
    V2
    V1 + V2
    Selected difference

    Equations and sign convention

    These are general complex-plane relations, not balancing formulas issued by ISO 21940. For each phasor with amplitude A and entered phase phi:

    x = A cos(phi)    ;    y = A sin(phi)
    V_sum = (x1 + x2) + j(y1 + y2)
    V_2-1 = (x2 – x1) + j(y2 – y1)    or    V_1-2 = -V_2-1
    A_result = hypot(x, y)    ;    phi_result = atan2(y, x)
    delta_phi_signed = wrap_to_[-180,180)(phi2 – phi1)

    A positive signed result means V2 is positive relative to V1 under the convention selected above: “lead” if positive angles are lead, “lag” if positive angles are lag. The smallest unsigned separation is the absolute value and intentionally discards direction.

    Zero magnitude has no phase. When an exact or floating-point-level cancellation makes the sum or difference magnitude zero, the worksheet reports phase as undefined. It does not display the arbitrary result of atan2 applied to numerical residue.

    Independent examples: 3 at 0 degrees plus 4 at 90 degrees gives 5 at 53.1301 degrees. For the same inputs, V2 minus V1 gives 5 at 126.8699 degrees. Equal opposing phasors have a zero sum with undefined phase; identical phasors have a zero difference with undefined phase.

    Balancing and diagnostic boundary

    The difference between a baseline response and a trial-run response can represent the measured response change only when the records are comparable. It is not by itself a correction weight. A balancing solution additionally needs the trial mass, radius, angular location, consistent measurement conventions, a valid influence coefficient and applicable safety/procedure controls.

    No “0 degrees = unbalance” fault ruleRelative phase depends on sensor direction, reference and transfer path. In-phase, out-of-phase or quadrature observations do not uniquely diagnose unbalance, misalignment or resonance. This worksheet makes no condition classification.

    Source and standards boundary

    Αριθμός ταυτότηταςΠηγήVerified public informationUse on this page
    S1ISO 21940-13:2012, Edition 1Published and confirmed in 2023; replaces ISO 20806:2009. It addresses criteria, instrumentation, safety, reporting and records for in-situ balancing of medium and large rotors. The public abstract explicitly says it does not provide methods for calculating correction masses from vibration data.Balancing safety/scope boundary only; not the source of the phasor equations.
    S2Dewesoft Order Analysis Manual V26-1Public manual pp. 14, 18-19 and 33-34: absolute order phase needs zero-angle information; phase reference may be the frequency-source/keyphasor position or the first order; extracted harmonics are complex channels with magnitude, phase, real and imaginary parts.Evidence for reference identity and complex-channel representation.
    S3Dewesoft Orbit Analysis Manual V26-1Public manual pp. 69-70 documents application-specific phase conventions: Order Analysis uses a leading zero-crossing convention while Orbit Analysis uses a lagging peak convention.Evidence that phase values from different modes are not interchangeable without conversion.
    S4Fluke, Phase in Vibration AnalysisOfficial technical article describes balancing with a rotational-rate response relative to tachometer, a known trial weight/angle and a subsequent program calculation.Public evidence that a response-vector difference alone is not a correction mass/angle method.
    NEEDS_LICENSED_SOURCEThe exact normative in-situ balancing conditions, instrumentation, safeguards, reporting and record requirements of ISO 21940-13 are protected. The official card verifies lifecycle and public scope only. No clause, tolerance, correction-mass method or acceptance criterion is invented here.

    Corrections made in this audit

    Signed and unsigned phase are no longer conflated

    The former page reduced every phase relation to 0-180 degrees, losing which phasor led or lagged. The replacement reports signed V2 minus V1 in [-180, 180) and the unsigned separation separately.

    Subtraction order is explicit

    V2 minus V1 and V1 minus V2 have the same amplitude but phases separated by 180 degrees. The user must now select which operation is intended, and the result/table label states it.

    Zero-vector phase is undefined

    The old engine called atan2 on floating residue for exact cancellation and could display a meaningless phase. The replacement uses a scale-aware numerical cancellation check and reports undefined phase.

    Measurement comparability is required

    Amplitude unit/scaling, coherent component, phase reference, sign/peak convention, sensor polarity, operating condition and measurement quality are now visible evidence gates.

    Balancing and diagnostic shortcuts were removed

    The old FAQ mapped 0/90/180-degree observations to specific faults and implied a simple phase shift supplied a correction angle. Those claims were removed; the page does not calculate a correction mass or diagnose a fault.

    Input and runtime integrity were rebuilt

    Defaults, presets, automatic calculation, browser history, clipboard export, dynamic HTML and external equation rendering were removed. Decimal point/comma are accepted separately; mixed separators, zero/negative amplitudes, non-finite/overflow inputs and incomplete records are rejected.

    Categories:

    WhatsApp
    Balanset-1A - €1975Ρωτήστε τον μηχανικό